DIGITEL Low Volume Sampler

for autonomous and continuous measurements

 

Model

DPA96

DPA02

 
Technical data
Type of construction Laboratory housing of anodised aluminium Laboratory housing of anodised aluminium
Power supply 230V +6/-10% / 50Hz / max.400VA 230V +6/-10% / 50Hz / max.50VA
Fuse protection 10A 10A
Specification connection cable 3 x 0,75 mm², 10 A, 250 V 3 x 0,75 mm², 10 A, 250 V
Installation Category II (Standard power) Category II (Standard power)
Field of application 5... 40 °C; 10 ... 90 % rF resp..
max. operation altitude: 2000 m above sea level.*
5... 40 °C; 10 ... 90 % rF resp..
max. operation altitude: 2000 m. above sea level*
Flow rate 0.2 - 2.5 l/min (DPA96*-02)
0.6 - 5.0 l/min (DPA96*-05)
1.0 - 10 l/min (DPA96*-10)
1.5 - 15 l/min (DPA96*-15)
2.0 - 25 l/min (DPA96*-25)
4.0 - 40 l/min (DPA96*-40)
5.0 - 50 l/min (DPA96*-50)
0,2 - 1,2 l/min
Volume flow“ control accuracy < 5% of MBE < 5% of MBE
Suction unit, average durability 36.000 h 12.000 h
reproducible accuracy settings (according to UMEG-report Nr. 6-08/00) +- 0,45% +- 0,45%
Accuracy logged standards and measured volumes: <+/- 2% <+/- 2%
Time programs Work, Pause ( 0 ... 59999 minutes each); Start time adjustable using date and time. Work (0 ... 59999 minutes each); Start time adjustable using date and time.
Outer dimensions 450 x 345 x 220 mm
(H x W x D)
370 x 236 x 154 mm
(H x W x D)
Weight ca. 20 kg ca. 5 kg
Protection class IP50 IP50
Underpressure

> 400 mbar at 50 l/min
> 500 mbar at 40 l/min
> 600 mbar at 30 l/min
> 700 mbar at 20 l/min

max. 200 mbar
Analogue remote control via interface
yes
yes
Logging via RS-232C
yes
yes
Switch-off overloading
yes
yes
motor loading display
yes
yes
Multi- lingual logging
yes
yes
Battery-backed data memory
yes
yes
Battery-backed clock module
yes (+/- 4 sec/ day)
yes (+/- 4 sec/ day)
Interface logging
DIGITEL, Bayern-Hessen-Protokoll, AK-Protokoll, customer specific
DIGITEL, Bayern-Hessen-Protokoll, AK-Protokoll, customer specific
Power consumption per day at a nozzle capacity of 75% and an activated pre-separator heating
approx. 7 kWh
approx. 1 kWh
Options
Inlets extern
TSP; PM10; PM2,5; PM1
TSP; PM10; PM2,5; PM1
Memory cards (PCMCIA) internal
yes
yes
Log printer
yes
yes
Ext. Meteo data logging
yes
yes
Ext. 16-port returnable valve
yes
yes
Ext. 8-port returnable valve
yes
yes
Field housing IP54
yes
yes

 

* Special construction for operation altitudes over 2000 m. above sea level on request

Dimension drawings:

DPA96 in Laboratory housing (pdf, ca. 82 kByte)
DPA02 in Laboratory housing (pdf, ca. 121kByte)


... top of page