|
DIGITEL High Volume Sampler Maintenance
Maintenance of the High Volume Samplers:
Digitel HVS DHA-80 requires minimum
maintenance. However, depending on the air pollution degree and climatic
load upon installation site, inspection of the sampler associated with
cleaning is necessary.
In particular, the following activities shall be performed:
Cleaning:
High-volume samplers must be cleaned on regular basis. Cleaning intervals
strongly depend on the installation site characteristics and they have
to be determined by the operator. They may range from one month up to
one year.
During cleaning, the instrument should be off power.
To clean the instrument, a dry cloth should be used. At heavy contamination,
the cloth should be wetted with a commercial window cleaning product.
Make sure that the instrument is dried up before starting up again.
Avoid using solving and scrubbing cleaning products!
The flow meter glass tube has to be visually inspected. In case of broken
filter or negligent sampler operation without a filter inserted, contamination
can also take place. In case of any doubt, the tube has to be removed
and cleaned.
Due to its difficult accessibility, the funnel-shaped flow chamber upper
part located before the filter, can only be cleaned in combination with
possible changer apparatus service works. As this section of air-sample
path shows a much larger inner diameter, it is less affected by deposits,
as a rule.
The inlet tube interior has to be inspected for wall deposits and in case
of doubt, cleaned, using a cloth. As a cleaning liquid, we recommend water
and/or spirits.
TSP sampling probes („open ring-slot“ according to VDE or
„EMPA/UBA“ - cylindrical probes) have to be checked for dust
deposits and cleaned, if possible. Normally cleaning using a wet cloth
is sufficient.
——————————————————————————————————————————————————————————————————————————————————————————————————
Exchange of sealing rings
The transition areas between the above-stated path separate section of
the air probes are equipped with sealing rings (O-rings).
Special attention has to be paid to the sealing ring of 44 x 3 mm at the
sampler inlet muffle, as well as to the glass measurement tube O-rings
(50.4 x 3.53 mm). These O-rings have to be checked and possibly replaced
after 2 to 3 years of operation.
The sealing ring of 150 x 3 mm at the bottom flange of upper part of the
flowing chamber should be inspected (by a DIGITEL service engineer) and
possibly replaced after 2 to 3 years of operation.
Sealing rings (150 x 3 mm) at the filter holder bottom part is coated
with an anti-friction layer. They have to be regularly checked when a
new filter is inserted and rubbed, using a dry cloth. When this layer
is worn out or in case of increasing sticking tendency, it should be renewed.
We recommend to replace these sealing rings annually.
——————————————————————————————————————————————————————————————————————————————————————————————————
Tightness test. Checking of volume-flow calibration
At start up, displayed blower load level, converter frequency, filter
type and flow will be recorded. Sudden insufficient blower capacity under
the same conditions, is caused by leakage in the air-sample path (after
the filter).
Another very simple option for testing the sampler tightness consists
in closing the sampler at the inlet muffle with the inlet tube removed
or by inserting an air-impenetrable cardboard instead of the filter paper
to the filter holder and switching the blower on. In both cases, the flow meter
floater must not be lifted from its resting position at the bottom of
the measurement tube. Hereby the blower must be run up to its maximum
capacity in order to reach the overload condition.
The checking of the volume flow simultaneously represents a check of tightness.
These procedures have to be taken approximately every two months.
The second flow meter of the same construction as used in the sampler itself,
controls the volume flow and is installed onto the sampler as „transfer
standard“ instead of a pre-separator or a suction pipe. With a new
round filter paper inserted, the positions of floaters are compared by
switching the blower on. At variations of the set point, originally calibrated
on the flow meter of the sampler, the checking of tightness should be performed.
——————————————————————————————————————————————————————————————————————————————————————————————————
Dusty round filter homogeneous deposit
Upon removing dusty filter papers or during weighing, filters have to
be subject to visual inspection for homogeneous deposits. Drop-like spots
in the filter centre, as a rule, indicate inoperable probe heating, or/and
a defective inlet muffle O-ring. Bright spots on the filter paper rim
are attributable to defective sealing of the flowing chamber upper part
with the filter holder upper surface (service works are definitely required!).
——————————————————————————————————————————————————————————————————————————————————————————————————
Blower
The blower has the average MTBF (average time between failures) of 36
000 hours. It is maintenance-free. However, for instruments under operation
for longer than two years, an occasional acoustic inspection of the blower
by opening the blower section, is recommended to prevent a possible blower
blocking.
Special attention has to be paid to excessive (unusual) noises generated
by the blower (scrubbing, screeching).
——————————————————————————————————————————————————————————————————————————————————————————————————
Filter preparation
Reliable and reproducible measurement results can only be achieved by
using filters that are carefully conditioned before and after sampling.
Filters are pre-weighed and provided with a date. In order to enable a
checking during the operation by which a correct assignment of filters
is possible, the filters are inserted into the filter holder, marked according
to respective dates. The spring collar is removed from the filter holder
(using pliers) while a Teflon ring is laid on a clean surface by using
forceps. New filters are removed from the filter magazine by using forceps
and laid into the filter holder. Then the Teflon ring should be laid again
(using forceps) on the ring of the spring collar by using pliers. Now
the filter holder is ready for transport to the sampler. During transporting
the filter, no impurities should get onto the filter (therefore refer
to standard EN 12341 Annex C).
The polluted filter is removed from the filter holder, using forceps,
and inserted into a simply folded parchment envelope.
Warning!
It should be noted that a possible labelling of a filter holder is only
permitted on its front, using a marker.
Any inscription on the filter holder on the upper or bottom sides, as
well as sticking labels (on the filter holder entire surface) might cause
problems with the filter exchange and is prohibited!
Please mind that no sealing ring (on the filter holder and in the flowing
chamber) get in touch with inscriptions. The solvents, applied in various
markers or pens, destroy the applied sealing rings! Moreover, paint residuals
may result in bonding the sealing rings!
——————————————————————————————————————————————————————————————————————————————————————————————————
Maintenance pre-separators:
To avoid effects of released separated rough dust particles, the
cannon surface of the impactor plate has to be permanently covered with
a thin fat layer. It has to be periodically renewed. As a grease, you
can use, for example: BAYSILON paste, high-vacuum grease, medium-viscous
(35 g tube) and silicon high-vacuum grease, medium Merck 100 g, Article
7922 LAB.
Thereby the life cycle depends upon the proportion of rough dust in the
sampled exterior air. It is recommended to clean the impactor plate after
14 sampling days, by the time the average total dust volume (TSP) on the
installation side is approx. 70 to 80 g/m3. With lower TSP, the cleaning
interval can be longer. You can extend the cleaning interval results by
rotating the moveable impactor plate, resting on the heating holder, by
approx. 15° (approx. 2 cm). Acceleration nozzles then point at the
„clean“ areas between rough dust deposit settled in a circular
form of the previous sampling operation.
The impactor plate can be removed simply after opening the probe upper
part . It has to be cleaned with a clean cloth and its cannon surface
has to be greased. An approx. 5 cm long band of grease should be equally
spread on the area, using a spatula. To relieve this maintenance in the
field, the impactor plate can be replaced by another plate prepared in
the laboratory.
Acceleration nozzles, probe casing liners, as well as a liner behind the
impactor plate with the above-mentioned TSP condition have to be cleaned
after 30 flowing days.
In case of longer sampling in foggy environment it is recommendable to
inspect the impactor plate for water condensate.
|